Permissible operating
conditions

+10 °C to +40 °C, ≤ 85 % relative air
humidity (non-condensing)

Permissible storage
conditions

- 20 °C to + 55 °C, ≤ 85 % relative air hu-
midity, 800 –1060 hPa ambient pressure

Power supply

4 x 1,5 V

AA alkaline batteries.

Battery life

For approx. 180 measurements,
depending on the blood pressure level
and/or pump pressure

Accessories

Instruction for use, 4 x 1.5 V AA alkaline
batteries., storage pouch

Classification

Internal supply, IPX0, no AP or APG,
continuous operation, type BF applied
part

Technical information is subject to change without notification
to allow for updates.

• This unit is in line with European Standard EN 60601-1-2 and

is subject to particular precautions with regard to electromag-
netic compatibility (EMC). Please note that portable and mobile
HF communication systems may interfere with this unit. More
details can be requested from the stated Customer Service ad-
dress or found at the end of the instructions for use.

• This device is in line with the EU Medical Devices Directive

93/42/EC, the “Medizinproduktegesetz” (German Medical De-
vices Act) and the standards EN 1060-1 (non-invasive sphyg-
momanometers, Part 1: General requirements), EN 1060-3
(non-invasive sphygmomanometers, Part 3: Supplemen-
tary requirements for electro-mechanical blood pressure
measuring systems) and IEC 80601-2-30 (Medical electrical

equipment – Part 2 – 30: Particular requirements for the safety
and essential performance of automated non-invasive blood
pressure monitors).

• The accuracy of this blood pressure monitor has been care-

fully checked and developed with regard to a long useful
life. If using the device for commercial medical purposes, it
must be regularly tested for accuracy by appropriate means.
Precise instructions for checking accuracy may be requested
from the service address.