Electromagnetic Compatibility Information

Table 1

Guidance and manufacture’s declaration – electromagnetic emissions

The BM 26 is intended for use in the electromagnetic environment specifi ed below.
The customer or the user of the BM 26 should assure that it is used in such an environment.

Emissions test

Compliance

Electromagnetic environment – guidance

RF emissions CISPR 11

Group 1

The BM 26 uses RF energy only for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in nearby electronic equipment.

RF emissions CISPR 11

Class B

The BM 26 is suitable for use in all establishments other than domestic and those
directly connected to the public low-voltage power supply network that supplies buil-
dings used for domestic purposes.

Harmonic emissions IEC 61000-3-2

Not applicable

Voltage fl uctuations/fl icker emissions
IEC 61000-3-3

Not applicable

Table 2

Guidance and manufacturer’s declaration – electromagnetic immunity

The BM 26 is intended for use in the electromagnetic environment specifi ed below. The customer or the user of the BM 26 should assure that it
is used in such an environment.

IMMUNITY test

IEC 60601 test level Compliance level Electromagnetic environment – guidance

Electrostatic discharge (ESD)
IEC 61000-4-2

± 6 kV contact
± 8 kV air

± 6 kV contact
± 8 kV air

Floors should be wood, concrete or ceramic tile. If fl oors are covered
with synthetic material, the relative humidity should be at least 30 %.

Power frequency (50/60 Hz)
magnetic fi eld IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fi elds should be at levels characteristic of
a typical location in a typical commercial or hospital environment.