Table 4

Recommended separation distances between portable and mobile RF communications equipment and the BM 70

The BM 70 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of
the BM 70 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the BM 70 as recommended below, according to the maximum output power of the communications equipment.

Rated maximum output

power of transmitter

(W)

Separation distance according to frequency of transmitter

(m)

150 kHz to 80 MHz

d = 1,2

80 MHz to 800 MHz

d = 1,2

800 MHz to 2,5 GHz

d = 2,3

0,01

0,12

0,12

0,23

0,1

0,38

0,38

0,73

1

1,2

1,2

2,3

10

3,8

3,8

7,3

100

12

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be deter-
mined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is aff ected by absorption and refl ection from struc-

tures, objects and people.