Guidance & Declaration — electromagnetic immunity

The TL90 is intended for use in the electromagnetic environment specified below. The customer or

the user of the TL90 should ensure that it is used in such an environment.

Immunity test

Iec 60601

test level

compliance

level

electromagnetic environment –

guidance

Radiated RF

IEC 61000-4-3

3 V/m

80 MHz to 2,5 GHz

3 V/m

Portable and mobile RF communications

equipment should be used no closer to any

part of the TL90, including cables, than the

recommended separation distance calcu-

lated from the equation applicable to the

frequency of the transmitter.

Recommended separation distance

3V

d=1.2 P

1/2

80 MHz to 800 MHz

d=2.3 P

1/2

800 MHz to 2.5 GHz

where P is the maximum output power rating

of the transmitter in watts (W) according

to the transmitter manufacturer and d is

the recommended separation distance in

metres (m).

Field strengths from fixed RF transmitters,

as determined by an electromagnetic site

survey,

a

should be less than the compliance

level in each frequency range.

b

Interference may occur in the vicinity of

equipment marked with the following

symbol:

 

NOTe 1 At 80 MHz and 800 MHz, the higher frequency range applies.

NOTe 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorpti-

on and reflection from structures, objects and people.

a

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land

mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically

with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagne-

tic site survey should be considered. If the measured field strength in the location in which the TL90 is used

exceeds the applicable RF compliance level above, the Medical Daylight should be observed to verify normal

operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting

or relocating the TL90.

b

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V/m.