4.0 Cleaning the Bluelab pH Probe

To ensure accurate readings the pH probe tip needs to be rinsed in water after each

use and cleaned prior to calibration using the following instructions.
The storage cap must always be put back on after cleaning. Always ensure it contains

enough Bluelab pH Probe KCl Storage Solution to cover the probe tip.

Remove storage cap from pH probe.
Hold the top of the storage cap, twist the

cap to loosen then remove.

Rinse pH probe tip under fresh tap water.
Never use RO (Reverse Osmosis), Distilled

or De-ionized water.

Fill a small plastic container with clean

tap water.
Add a small amount of Bluelab pH Probe

Cleaner or mild detergent (dishwashing liquid).

Gently stir the probe tip in the mixture.
Ensure that you do not ‘knock’ the pH

probe on the side of the container as this

may cause damage to the probe.
Rinse well under fresh running water to

remove all traces of the detergent mixture.

If the probe tip requires removal of heavy

contamination:
Gently brush around the glassware with a

few drops of Bluelab pH Probe Cleaner or

mild detergent (dishwashing liquid) and a

soft toothbrush.

Rinse well under fresh running tap water to

remove all traces of the detergent mixture.

Calibrate pH probe to meter after cleaning.
Follow the calibration instructions on the

back of the meter, in the manual or watch

our videos online.

5.0 Storing the Bluelab pH Probe

To prepare the pH probe for storage:
Add enough Bluelab pH Probe KCl Storage

Solution into the probe storage cap to fully

submerge the pH probe tip. Then replace

the cap and store in a secure place.
DO NOT use RO (Reverse Osmosis), Distilled

or De-ionized water. Pure water changes

the chemistry in the reference, causing the

probe to die.