Troubleshooting guide, 0 troubleshooting guide – Bluelab pH Probe User Manual

Page 7

Advertising
background image

English

7

8.0 Troubleshooting

guide

Trouble

Reason

Correction

pH reading does not

change from solution

to solution

Broken glass bulb,

tube or connector

Check pH probe for damage.

Replace probe.

pH reading inaccurate
(Drift, readings slowly

varying)

Contaminated pH probe /

glassware not clean

Clean pH probe (see section 4.0);

then calibrate.

Wick contaminated,

blocked or dry

Hydrate probe in KCl storage

solution for up to 24 hours, see

section 6.0. Avoid measuring

proteins or oils. Replace unit.

Incorrect pH calibration

Ensure calibration solutions are

accurate. Replace if in doubt.
Wait longer for readings to

stabilize to a constant value

before calibrating.
Calibrate using two points - 7.0 pH,

then 4.0 pH or 10.0 pH.

pH calibration unreliable

Calibrate pH probe.

pH probe damaged or old. Replace pH probe.

Incorrect sample

reading following

successful calibration

Ground loop (often occurs

in process systems)

Verify by removing the sample

from its environment then take

measurement. May require electrical

circuitry to be checked in system.

Wick blocked

Replace probe.

Unsuccessful

calibration

(Low slope <90%)

Buffers inaccurate

Replace buffers.

Glassware not clean

Clean pH probe (see section 4.0);

then calibrate.

Glassware aged

(glassware will not clean)

Replace probe.

BNC fi tting wet

Use cloth to dry (note terms of

guarantee).

Noisy - readings

jumping

Poor connection to meter Check probe is connected to

meter correctly.

Contact zone not

immersed

Lower probe into solution at least

2 cm / 1”.

Displays pH 7

for all buffers

Electrical short

Check BNC fi tting and cable for

damage.

BNC fi tting wet

Use cloth to dry (note terms of

guarantee).

Broken or cracked glass

bulb or tube

Replace probe.

Advertising