6

6.0 Hydrating the pH probe

Hydrate the pH probe in Bluelab pH Probe KCl Storage Solution when:
• the probe tip has not always been stored in KCl storage solution, to improve the

reading response speed.
• the probe tip has been accidentally allowed to dry out.
Never use RO (Reverse Osmosis), De-ionized or Distilled water.

Pure water changes the chemistry in the reference, causing the probe to die.

Loosen, then remove the storage cap.

Place the pH probe upright in a plastic

container.

Clean the pH probe tip.

Ensure the probe tip has been cleaned

before hydrating. See section 4.0 for

instructions.

Add enough Bluelab pH Probe KCl

Storage Solution to a plastic container to

submerge the pH probe tip.

Leave to soak for at least 24 hours.

After hydration, always calibrate the pH

probe to ensure accuracy.

1

2

3

4

Bluelab pH Probe KCl
Storage Solution

7.0 Technical

specifi cations

pH

Measurement range

0.0 - 14.0 pH

Accuracy at 25 °C/77 °F

±0.1 pH

Reference

Double junction, Ag/AgCl, Sat. KCl gel

Operating environment

0 - 60 °C

32 - 144 °F