N 

10 

Parameter index 16 (return)

Parameter number 31

Capacity unit 

21 

01 

00000001  P

01 

Process 1

22 

7F 

01111111 

T 

60 

Parameter type ‘string’

P

1F 

Parameter number (FBnr.) 31 – Capacity unit

23 

07 

00001110   

07 

String length 7

24 

71 

01110001 

C

00 

Parameter not chained (return)

Parameter index 17 (return)

Process 1 

Parameter number 17 

Fluid name 

T 

60 

Parameter type ‘string’ (return)

N 

11 

Parameter index 17 (return)

25 

01 

00000001  P

01 

Process 1

26 

71 

01110001 

T 

60 

Parameter type ‘string’

P

11 

Parameter number (FBnr.) 17 – Fluid name

27 

0A 

 

 

 

String length 10

29 

\n   

 

 

Line Feed

 

28 

\r 

 

 

 

Carriage Return

End

Total String 

:1A0304F1EC7163006D71660001AE0120CF014DF0017F077101710A\r\n 

 

Answer by node 3

Nr 

Byte  Layout 

Description 

Description per block

0 

: 

 

 

Start 

1 

37 

 

Number of bytes which do follow: 65 bytes

2 

03 

 

Node 3 

3 

02 

 

Command write

4 

F1 

11110001 

C 

80 

Process chained

Process 113 

P 

71 

Process 113 (receiving process)

5 

EC 

11101100 

C 

80 

Parameter chained

Parameter index 12 

Serial number 

T 

60 

Parameter type ‘string’

N 

0C  Parameter index 12

6 

00 

 

Length of the answer 10 Bytes

7‐26 

 

 

4D 36 32 31 32 33 34 35 41 00
Parameter value converted from hex to ASCII : M6212345A 

27 

6D 

01101101 

C 

00  Process not chained

Parameter index 13 

USERTAG 

T 

60  Parameter type ‘string’

N 

0D  Parameter index 13

28 

00 

 

String length 00, length not defined

29‐36 

 

 

55 53 45 52 54 41 47 00
Parameter value converted from hex to ASCII, the values do 
read : USERTAG 

37 

01 

00000001 

C 

00  Process not chained

Process 1 

P 

01  Process 1 (receiving process)

38 

AE 

10101110 

C 

80  Parameter chained

Parameter index 14 

Measure 

T 

20  Parameter type ’integer’

N 

0E  Parameter index 14

39 

1C 

 

Parameter value is: 1CD8 (hex)
Measure Value is:  7384 (dec) 

 

40 

D8 

41 

CF 

11001111 

C 

80  Parameter chained

Parameter index 15 

Capacity 

T 

40  Parameter type ‘float’

N 

0F 

Parameter index 15

42 

3F 

 

3F 80 00 00 
Parameter Value in IEEE‐floating point notation, 32‐bit single 
precision  
Parameter value converted from float to decimal, the values 
reads : 1.0 

43 

80 

44 

00 

45 

00 