3 plumbing, 1 sample requirements, 2 connecting the inlet and drain fittings – Electro-Chemical Devices (ECD) SMS22 User Manual

Page 15: 3 connecting the reagents

Advertising
background image

2.3 PLUMBING

2.3.1 Sample Requirements
Minimum flow: 0.4 gal/hour, (1.5 L/hr)

Temperature: 32° to 122°F (0° to 50°C)

Sample Pressure: Drawn from atmospheric pressure

The Sample Conditioning Cell provides an atmospheric sample point that can adapt to changing sample

flows between 0.4 and 10 gal/hr (1.5 -37 L/hr).

2.3.2 Connecting the Inlet and Drain fittings
Sample Inlet:

The sample must be drawn from atmospheric pressure. The sample feed tube is 8 mm OD x 4.8 mm

bore (5/16” OD x 3/16” bore). The sample tube can be placed directly in an open reservoir of the sample

water or connected to the reservoir with a 3/16” barb fitting.
A Sample Conditioning Cell is available for pressurized samples. This Fast Flow Reservoir allows the

sample to fill a 200 ml sampling cell and overflow the balance of the feed to an atmospheric drain. The

input to the sample cell is a ¼” barb fitting is supplied for sample

connection but a compression style tube fitting can also be used (not

supplied). The sampling port is a stainless steel tube that

accommodates the sample feed tubing to the analyzer. The overflow

drains through the 12 mm overflow tubing exiting the bottom of the

Sample Conditioning Cell. Attach a length of 12 mm soft tubing and

allow the overflow to drain to open atmosphere. Do not restrict the

drain line. The pressurized feed must have an adjustable shut off

valve. Adjust the feed so that the Sample Conditioning Cell fills and

starts overflowing in 2 minutes or less. Filling the cell in 2 minutes

provides the minimum sample flow (1.5 liters/hour), faster feed rates

simply overflow to the drain.
Sample Drain:

The neutralized sample and the water from the rinse cycles drain

from the measurement cell when the solenoid valve is actuated. The

drain line must drain to atmospheric pressure.

2.3.3 Connecting the Reagents

DANGER: Read any precautions and the (MSDS) datasheets, wear protective gloves,

clothes and glasses before handling chemical products.

Hazard of severe burns or injury due to handling of dangerous chemicals

Connect the tubing from the reagent pumps to the insertion tubes. Press the insertion tube from
Reagent Pump 1 through the grommet in the cap of the 16% KOH solution. Press the insertion tube from
Reagent Pump 2 through the grommet in the cap of the 7.5% HCl solution.

SMS-22

Page 6

Advertising