Standard setting – Flowserve NRGS 11-1 User Manual

Page 14

Advertising
background image

14

The measuring range can be switch selected between

0.5 µS/cm and

10 µS/cm

by means of code switch :

1. Unscrew screws at the electrode terminal, remove cover , Fig. 6

DIP switch 1 – 4 OFF

Measuring range

0.5 µ S/cm.

DIP switch 1 – 4 ON

Measuring range

10 µ S/cm.

2. Replace cover and fasten screws .

Standard setting

Factory setting

The compact system is delivered with the following factory settings:

Measuring range

10

µ

S/cm

Switch selection of measuring range

Attention

Screwdriver for cross head screws, size 1

Screwdriver for slotted screws, size 2.5; completely insulated according
to VDE 0680

Do not damage the electronic components when setting the code
switch.

Tools

5

E

G

E

G