General information, 1 introduction, 2 application – Fluke Biomedical 6000-529 User Manual

Page 5: 3 specifications

Advertising
background image

General Information

Introduction

1

1-1

Section 1

General Information

1.1 Introduction

The Model 6000-529 Mammographic Ion Chamber is an ion chamber especially designed to measure low

energy x

-

ray exposure as encountered in mammography.

1.2 Application

The Model 6000-529 Mammographic Ion Chamber is specifically designed for measurement of exposure
rate from the output of mammographic x-ray tubes.

1.3 Specifications

Volume 3.3

cm

3

Size

41 mm x 14 mm high

Window Material

Polycarbonate

Window Thickness

9.5 mg/cm

2

Sensitivity

1 nC/R, nominal (0.087 nC/Gy)

Energy Dependence

Within ± 5% from 0.2 mm AI HVL to 5.0 mm AI

Angular Response

± 2% for radiation incidence of up to 30° from normal

Max. Exposure Rate

13 R/min at 300 V bias for 99% collection efficiency

Accessories

Custom carrying case and detachable acrylic stem

Cable Length

10 meters

Cable Termination

BNC signal and banana HV or BNC Triax

Weight

128 g