Fluke Biomedical ESA615 User Manual

Page 18

Advertising
background image

ESA615
Users Manual

4

•

Do not use test leads if they are
damaged. Examine the test leads for
damaged insulation, exposed metal, or if
the wear indicator shows. Check test
lead continuity.

•

Examine the case before you use the
Product. Look for cracks or missing
plastic. Carefully look at the insulation
around the terminals.

•

Make sure the ground conductor in the
mains power cord is connected to a
protective earth ground. Disruption of
the protective earth could put voltage on
the chassis that could cause death.

•

Do not use an extension cord or adapter
plug.

•

Do not use and disable the Product if it
is damaged.

•

Do not use the Product around explosive
gas, vapor, or in damp or wet
environments.

•

Do not touch voltages > 30 V ac rms, 42
V ac peak, or 60 V dc.

•

Measure a known voltage first to make
sure that the Product operates correctly.

•

Do not apply more than the rated
voltage, between the terminals or
between each terminal and earth ground.

•

Remove the null post adapter from the

∅

/Null jack after a test lead zero is

performed. The

∅

/Null jack becomes

potentially hazardous during some of the
test conditions. Use only cables with
correct voltage ratings.

•

Remove all probes, test leads, and
accessories that are not necessary for
the measurement.

•

Keep fingers behind the finger guards on
the probes.

•

Connect the common test lead before
the live test lead and remove the live test
lead before the common test lead.

•

Use only current probes, test leads, and
adapters supplied with the Product.

•

Do not use the 15-20 A adapter to supply
power to devices rated more than 15 A.
This can overload the installation.