Electrical Safety Analyzer

How to Do Electrical Safety Tests

49

CO

N

DUCTI

V

E PART

DUT_PE

DUT_L2

DUT_L1

FE

DE

V

ICE U

N

DER TEST

APPLIED

PART

LEAD

SELECT

RELAY*

*Leads not selected
are open.

TEST
LEAD

MD

gtv23.eps

Figure 28. Alternative Applied Part Leakage Test Schematic