Operation, 1 operation – Fluke Biomedical 956A-201-S1 User Manual

Page 11

Advertising
background image

Operation

Operation

4

4-1

Section 4

Operation

4.1 Operation

The UDR Simulator displays an mR/h value based on input voltage. Alarms and analog outputs operate
as they would in a normal, area radiation-monitoring channel. High and Alert alarms are factory
configured to require manual, or operator, reset. If automatic reset of the High and Alert alarms is desired,
configuration jumper JP3-3 must be removed. The Table below shows the expected UDR Simulator
radiation display values for input voltages based on a conversion constant of 6.25-1.

Signal Converter

Input

Signal Converter Output/ UDR Input

UDR Simulator Display

-10 VDC

0 VDC

0.01 mR/h nominal

- 6 VDC

2 VDC

0.10 mR/h

- 2 VDC

4 VDC

1.0 mR/h

+2 VDC

6 VDC

10 mR/h

+6 VDC

8 VDC

0.10 R/h

+10 VDC

10 VDC

1.0R/hr to EEEEE *

* Over range is 1.00E3 mR/h

Fine span adjustment of the UDR Simulator display is accomplished through potentiometer R1 on the
942-200-75 Analog Input Option circuit card. The Signal Isolator is also provided with zero and span
adjustment potentiometers.

Refer to the 955A Standard Manual for UDR display features, alarm function/logic and operator
interfaces.