Receiving inspection and storage, 1 receiving inspection, 2 storage – Fluke Biomedical 956A-201-S1 User Manual

Page 7

Advertising
background image

Receiving Inspection and Storage

Receiving Inspection

2

2-1

Section 2

Receiving Inspection and Storage

2.1 Receiving Inspection

Upon receipt of the unit:

1. Inspect the carton(s) and contents for damage. If damage is evident, file a claim with the carrier

and notify Fluke Biomedical, Radiation Management Services at 440.248.9300.

2. Check that all items on the packing slip are present and in good condition. If any items are missing

or damaged, contact Fluke Biomedical at 440.248.9300.

2.2 Storage

Instruments storage must comply with Level B storage requirements as outlined in ANSI N45.2.2 (1972)
Section 6.1.2(.2). The storage area shall comply with ANSI N45.2.2 (1972) Section 6.2 Storage Area,
Paragraphs 6.2.1 through 6.2.5. Housekeeping shall conform to ANSI N45.2.3 (1972).

Level B components shall be stored within a fire resistant, tear resistant, weather tight enclosure, in a
well-ventilated building or equivalent.

Instruments storage must comply with the following:

1. Inspection and examination of items in storage must be in accordance with ANSI N45.2.2 (1972)

Section 6.4.1.

2. Requirements for proper storage must be documented and written procedures or instructions must

be established.

3. In the event of fire, post-fire evaluation must be in accordance with ANSI N45.2.2 (1972), Section

6.4.3.

4. Removal of items from storage must be in accordance with ANSI N45.2.2 (1972), Sections 6.5 and

Advertising