Section 1, Introduction, 1 general description – Fluke Biomedical 960CI-200 User Manual

Page 5: 2 application, 3 specifications

Advertising
background image

Introduction

1

1-1

Section 1

Introduction


1.1 General Description

The 960CI-200 series Communications Isolator is designed for use in Victoreen’s 960 Digital Radiation Monitoring
System for Class 1E applications. Its function is to isolate the safety related monitor from a commercially rated
computer both electrically and mechanically. The monitor transmits data serially to the 960CI isolator through a fiber
optic cable. Figure 1-1 is a block diagram of the Communications Isolator Module.

1.2 Application


The 960CI-200 series Communications Isolator Module is used for communication isolation whenever a safety related
monitor needs to be connected to a computer. This isolation is for the serial asynchronous communication link. The
960CI is designed for use with the 960IC Card Cage Assembly, which holds up to 8 960CI modules.

1.3 Specifications

General specifications for the 960CI-200 Communications Isolator Module are listed below. The module (960CI-200)
is only available rated for nuclear safety-related applications. However, the module may also be used in commercial
applications.

The 960CI-200 module is assembled by techniques and with parts selected for the reliability required in a nuclear
application. Any repairs made to the nuclear rated module will void its safety-related rating. The module (960CI-200)
must be returned to the factory for authorized qualified ANSI 45.2.6, 1978, Skill
Level II service.

Specifications:

Dimensions (H x W x D)

12in x 6.5in x 1in (304.8mm x 156.2mm x 25.4mm)

Weight

1

lb.,

9

oz.

(708.74g)

Operating

Temperature

32

° to 122°F (0° to 50°C)

Relative Humidity

0 to 95% noncondensing

Power

+

5

VDC

@

500mA,

115

VAC

Isolation

Optical, 50 kV minimum

Processor

Motorola

6809

Data Bus

8 internal data lines D0 through D7

Address Bus

16 internal address lines A0 through A15

Control & Timing

01, 02 two phase 1 MHz clock R/W, short 02 early 0 for

WRITE reset

Channel ID

32 codes, switch selectable

Communication

One way optical serial line to a safety related monitor;

two way serial link to a computer

Baud

Rates

See

Table

4-2

Memory

32K x 8 PROM, 24 x 8 RAM




Advertising