2 specifications – Fluke Biomedical 05-434 User Manual

Page 6

Advertising
background image

Nuclear Associates 05-434
Operators Manual

1-2

1.2 Specifications

Visual Alarm

Two bright red incandescent lamps which flash at the same rate and at the
same time as the Primalert to which it is connected.

Audible Alarm

400 Hz tone alternating and actuated as on Primalert.


Alarm Switch

Permits the operator to disable the Audible Alarm.


Power Indicator

Green LED indicates power is being supplied to the Primalarm.


Fail-Safe Features

Three items will cause the Primalarm to alarm continuously (no on-off

cycle):

• The Primalert to which it is interfaced is not connected to primary power.
• The cable connecting the Primalert to the Primalarm is broken or

disconnected.

• The internal power or power converter on the PRIMALERT’” is defective.


Power

Part 14-314 (United States, Canada)

Input 117 VAC, 60 Hz, 12 W

Output 12 VDC, 500 mA

Part 14-400 (Europe)

Input 230 VAC, 50 Hz,

Output 12 VDC, 500 mA

Part 14-417 (United Kingdom)

Input 230 VAC, 50 Hz, 13.8 W

Output 12 VDC, 580 mA

Part 14-417, Part 14-416 (Australia)

Dimensions

5.9 in. x 3 4 in x 1.3 in. (15 cm x 8.6 cm x 3.3 cm)

(H x W x D)

Unit clips to wall with convenient mounting clip


Operating Conditions

10° C to 40° C (50° F to 104° F)

Maximum of 90% relative humidity non-condensing


Accessories Supplied

y 100 ft. (304.8 mm) cable with plugs on both ends to connect unit to

Primalert.

y Wall mounting bracket P/N 102007
y Appropriate 115V (230) CE approved converter

Advertising