4 making a measurement – Fluke Biomedical 37-720 User Manual

Page 10

Advertising
background image

37-720
User Manual

2-2


If two detectors are connected, alternate readings
can be taken by switching between DET A and
DET B. Both detectors should be calibrated
concurrently.

The highest degree of calibration can be attained
by calibrating at a dose or dose rate near ½ scale
(1000 cGy or 1000 cGy/min). This is due to display
resolution.


After the battery check, instrument zeroing, and calibration have been performed, measurements can be
taken.

2.4 Making a Measurement

Rate: cGy/min

1. Zero the instrument if not previously zeroed. Refer to "Zeroing the Instrument", at the beginning of

Section 2.2.

2. Connect the appropriate probe/probes to the DET A, DET B inputs.

3. Select the RATE mode on the front panel.

4. Initiate a radiation exposure.

5. During the exposure, view the reading on the display.

Dose Mode: cGy

1. Zero the instrument if not previously zeroed. Refer to "Zeroing the Instrument", at the beginning of

Section 2.2.

2. Connect the appropriate probe/probes to the DET A, DET B inputs.

3. Select the DOSE mode on the front panel.

4. Initiate a radiation exposure.

5. Upon completion of the exposure, view the results on the front panel display.

6. Alternate between DET A and DET B to view both channels.

NOTE