Operation

Source Geometry Corrections

2

2-9

Range Position: 100 mR/h

Indication (On “0-10” meter arc): Four (4) major divisions plus two and one-half (2½) minor divisions.

Major
Divisions

Minor

Divisions

Reading: 4

10

+

2.5
50

40

100

+

5
100

=

45
100

45
100

X

100 mR = 45 mr/h





















Figure 2-4

Reading Meter Scale Indications