Victoreen 440RF/D
Operators Manual

1-2

1.2 Specifications

Full Scale Range

Model 440RF/D:
0 to 1, 0 to 3, 0 to10, 0 to 30 and 0 to 100 mR/h
Model 440RF/D-SI
0 to 10, 0 to 30 and 0 to 100, 0 to 300 and 0 to 1000 μSv/h

Radiation Detected

Gamma, x-ray, and beta

Operating Range

Model 440RF/D: Exposure rate in five overlapping ranges:
0 to 1, 0 to 3, 0 to 10, 0 to 30 and 0 to 100 mR/h
Model 440RF/D-SI: Dose equivalent in five overlapping
ranges: 0 to 10, 0 to 30, 0 to 100, 0 to 300, 0 to 1000 μSv/h

Energy Response

X and Gamma: + 10% from 12.5 keV to 42 keV effective, with
a maximum peak of + 40% in the medium energy range.
Beta: Energies above .15 MeV

Insensitivity RFI:

Operates properly in 10 mW/cm

2

fields

Magnetic: Operates properly in a 200 gauss field

Accuracy

Maximum instrument inaccuracy, exclusive of energy
dependence, is less than + 10% of full-scale indication.

Response Time

Range Response

Time

0 to 1 mR/h (0 to 10 μSv/h)

7 seconds

0 to 3 mR/h (0 to 30 μSv/h)

7 seconds

0 to 10 mR/h (0 to 100 μSv/h)

5 seconds

0 to 30 mR/h (0 to 300 μSv/h)

5 seconds

0 to 100 mR/h (0 to 1000 μSv/h)

5 seconds

Ion Chamber

Chamber Window Area: 10 cm

2

Volume: 60 cc
Chamber Window: Approximately 1.5 mg/cm

2

aluminized

mylar
Dimensions: 2.5 in (6.4 cm) long; 1.5 in (3.8 cm) diameter

Case Window

13 mg/cm

2

magnesium

Batteries

Three 9 V alkaline transistor batteries for the electronics (9 V)
and two 9 V alkaline transistor batteries for the chamber (-18
V) provide over 200 hours of continuous operation.

Weight

6 lb, 6.5 oz (2.9 kg)

Dimensions (H x W x L)

Unit: 8 in (includes handle) x 4 7/8 in x 10 7/8 in
(20.3 cm x 10.4 cm x 25.7 cm)
Display: 3.7 in (9.5 cm) Meter scale, marked 0 to 3 and 0 to
10

Environmental Dependence

Operating Temperature: -4° F to 104° F (-20° C to 40° C)
Relative Humidity: 0 to 95% non-condensing