Mca527 – GBS Elektronik MCA-166 Communication DLL User Manual

Page 16

Advertising
background image

MCA527

Name

MMCA_SET_FLAT_TOP_TIME

Description

The function sets the flattop time.

Declaration

ERROR_FLAG CALLBACK MMCA_SET_FLAT_TOP_TIME(unsigned short unFlatTopTime);

unFlatTopTime

Flattop time:

0 … 50 * 0.1 µs

Return value

See MCA_COMM function

Remarks

This function is only usable for MCA527.

Name

MMCA_SET_JITTER_CORRECTION

Description

The function sets the jitter correction.

Declaration

ERROR_FLAG CALLBACK MMCA_SET_JITTER_CORRECTION(unsigned short unOffOn);

unOffOn

0 turns jitter correction off and

0 turns it on

Return value

See MCA_COMM function

Remarks

This function is only usable for MCA527.

Name

MMCA_SET_BASELINE_RESTORING

Description

The function sets the baseline restorer.

Declaration

ERROR_FLAG CALLBACK MMCA_SET_BASELINE_RESTORING(unsigned char uchOption);

uchOption

See CMD_SET_BASELINE_RESTORING

Return value

See MCA_COMM function

Remarks

This function is only usable for MCA527.

Name

MMCA_SET_PILE_UP_REJECTION

Description

The function sets the pile up rejection.

Declaration

ERROR_FLAG CALLBACK MMCA_SET_PILE_UP_REJECTION(unsigned short unOffOn);

unOffOn

0 turns PUR off and

0 turns it on

Return value

See MCA_COMM function

Name

MMCA_SET_FAST

Description

The function sets the fast discriminator threshold.

Declaration

ERROR_FLAG CALLBACK MMCA_SET_FAST(unsigned short unFast);

unFast

0 ... 2499, default 400

Return value

See MCA_COMM function

Name

MMCA_SET_SLOW

Description

The function sets the slow discriminator threshold.

Declaration

ERROR_FLAG CALLBACK MMCA_SET_SLOW(unsigned short unSlow);

unSlow

0 ... 2499, default 400

Return value

See MCA_COMM function

16