Description of the MCA Communication DLL

3 Sending Firmware Commands to the MCA

For information about MCA firmware commands (CMD_

XXX

) see document “Description of the

MCA527 Firmware Commands” or appendix “Description of the MCA166 firmware functions” within
the MCA-166 User's Manual

.

Name

MCA_COMM, MCA_COMM6

Description

These functions allow to send single firmware commands to the MCA. These function are only intended
for testing. The function MCA_COMM6 has been added because new MCA527 commands need more
than 4 parameters. For the real work with the MCA you should use the other functions (MMCA_ ...)
described in chapter 4.
Regardless of its type, each parameter of the firmware command is passed to one unsigned long
parameter each. It will be cast within the function to the required type.

Declaration

ERROR_FLAG CALLBACK MCA_COMM(LPSTR command,LPSTR rec_data,

unsigned long param1=0,unsigned long param2=0,
unsigned long param3=0,unsigned long param4=0);
ERROR_FLAG CALLBACK MCA_COMM6(LPSTR command,LPSTR rec_data,

unsigned long param1=0,unsigned long param2=0,
unsigned long param3=0,unsigned long param4=0

unsigned long param5=0,unsigned long param6=0);

command

Command name (e. g. “CMD_START”)

rec_data

Pointer to RECDATA or command specific structure in which the received data is returned.

param1

1

st

parameter if required

param2

2

nd

parameter if required

param3

3

rd

parameter if required

param4

4

th

parameter if required

param5

5

th

parameter if required

param6

6

th

parameter if required

Return value

ERROR_OK

= 0

successful data transfer

ERROR_INTERFACE

= 1

communication port is not initialized

ERROR_UNKNOWN_COMMAND

= 2

unknown command

ERROR_COMMUNICATION

= 3

faulty data transfer

ERROR_INVALID_PARAM

= 4

invalid parameter

ERROR_RUNNING_MEAS

= 5

measuring is running, but stopped
measurement is required for this command

ERROR_VIOLATED_RIGHT

= 6

execution right violation

ERROR_STOPPED_MEAS

= 7

measurement is stopped, but running
measurement is required for this command

ERROR_WRONG_MODE

= 8

wrong mode for using this command

ERROR_UNHANDLED_COMMAND

= 9

not handled by this firmware version

ERROR_FILE_WRITING_IN_PROCCESS

= 10

file writing is in process; this command must
not be called before the process is finished

7