Te70xp and te77xp – Hoefer TE70X User Manual

Page 18

Advertising
background image

p10

TE70XP and TE77XP

1

Turn the instrument on using the POWER button on
the front keypad.
The instrument works by setting the current and time,
and then starting the transfer. During the transfer, the
voltage can be displayed. The transfer cannot be set
to run at constant voltage.
The DISPLAY MODE BUTTON toggles between the
current (mA), time (hour:minutes) and the voltage (V).
The UP and DOWN arrows to change the value for the
current and time.

2

Press the START/STOP button to start the transfer.

Variable

Units

Range

Increment

Current

milliamps

1– 500

1 mA

Time

hours:minutes

5 min – 4:00 hours

5 min

Voltage

volts

0 – 30 (read only)

1 volt

Note: While setting the time,
there is an extra setting labelled
“ON” above the 4:00 upper
limit. This will disable the
timer, allowing the transfer to
run continuously until manually
turned off by the user.

transfer
start/stop

up/down arrows
adjust value

display mode
cycles LED display

power
ON/OFF

Fig 6. TE70XP keypad.

Advertising