Hoefer TE70X User Manual

Page 20

Advertising
background image

•

p12

After the transfer is complete

TE70X and TE77X
Turn off the power supply and disconnect the
leads from the power supply.

TE70XP and TE77XP
After a transfer is complete, the unit will beep
for 5 seconds. The LED display will blink and
cycle between the values for the current, time
and voltage at the time the transfer was stopped.
These can be recorded into a notebook, if desired.
Pushing any button will clear the display.

1

Remove the cover slowly because the stack may
adhere to it.

2

Remove and dispose of the upper blot papers.

3

Remove the gel(s).
Optional: Stain gel to check for residual protein left
in the gel.

4

If desired, label the gel contact side of the membrane
with a soft pencil.

5

Remove the membrane(s) from the stack with
blunt forceps. Process the membrane according to
your protocol or allow the membrane to air dry prior
to storage.

6

Remove the remaining blot papers and dispose.

7

Rinse the unit according to the Care and maintenance
instructions (next page).

Note: Staining the gel(s) for
residual protein gives an indication
of the completeness of transfer.

Note: Rewet dried membranes in
the appropriate wetting buffer
prior to processing.