5 calibration – Hoefer DQ300 User Manual

Page 17

Advertising
background image

p9

3.5 Calibration

1

Press the <CAL> button.

2

Press <ENTER> to start the calibration.

3

Insert your blank and press <ENTER>. The DQ300
will average the fluorescence of the blank for
5 seconds.

4

Insert the calibration standard and press <ENTER>.
The DQ300 will average the fluorescence of the
standard for 5 seconds.

5

Press <ENTER> when the calibration is complete to
accept the calibration. If <ENTER> is not pressed
within 10 seconds, you will be asked if you want to
abort the calibration. Press the <UP> or <DOWN>
button to abort or accept the calibration, respectively.


If at any time during steps 1–4 you want to stop
the calibration, press <ESC>. The display will
return to the Home Screen and default back to
the previous calibration.

Note: If using the Minicell
adaptor, be sure to insert it in
the correct position (see 1.3.5
on page 4).