4.1. PROTECTIVE OPERATIONS/OPERATION

DISTURBANCES

The Procool 10 unit is protected against dry operation with pressure guard (less than 1 bar)

which is on pressure side and against overheating (more than 65 °C) with sensor which is

placed into reservoir.

Disturbance situation

Pilot lamp H2 for pressure illuminates

during welding, welding and liquid circulation

are stopped

Pilot lamp H1 for temperature illuminates,

welding is stopped, but liquid circulation

is continued.

Pilot lamp H2 is flashing when there is normal

pause in liquid circulation

Reason

In system low liquid pressure during welding

Locate liquid leakage or operating disturbance

with switch S2. Check liquid amount.

Liquid overheating.

Wait until pilot lamp H1 is switched off.

Pressure guard is broken or there is pressure in

system.

Return hose of current

surface

level

Temperature sensor

Adjustment screw

Pressure guard

NOTE! If pilot lamp for pressure illuminates, liquid circulation operates from test switch S2.

Isn’t there any leakage, check disturbance as follows:
1.

Press on switch S2.

2.

Check through filling opening of reservoir that liquid is returned to reservoir. If any

liquid doesn’t come, take contact with nearest KEMPPI authorised service repair shop.

3.

Continue pressing on switch S2 and adjust freely adjustment screw of pressure guard

in such a way that pilot lamp H2 is only just switched off.

4.

Release switch S2 and wait for approx. 10 s and check that pilot lamp H2 is not

flashing.

5.

Still check by welding for approx. 20 s that there aren’t any more disturbances.
Daily check with test switch that liquid is also returned to reservoir.

4.2. CONTROL FUSE

On circuit card there is the fuse FA101 of 0,63 A, slow-blow.

A use same type and rating of fuse which is marked beside the fuse adapter

Damage caused by a wrong type fuse, is not covered by the guarantee.