Interface cable requirements, Customer interface cable requirements – MicroE Mercury II 6000V User Manual

Page 18

Advertising
background image

Section C - Reference

Page 18

Customer Interface

Cable Requirements

Mercury

II

6000

Signal

Twisted Pair

A+

Pair 1

A-

B+

Pair 2

B-

Index+

Pair 3

Index-

Left Limit +

Pair 4

Left Limit -

Right Limit +

Pair 5

Right Limit -

+5V

Pair 6

GND

C4.0

Customer cables that interface to Mercury

II

series

encoders must have the following characteristics:

• Twisted pair signal wiring.

• Characteristic impedance of 100-120 ohms.

• Sufficient wire gauge to meet the minimum

voltage requirement at the encoder, for example
24AWG gauge wire for a 2m length cable.
Examples of acceptable cables with 24AWG
gauge wire and 4 twisted pairs are Belden
9831, 8104, and 9844 or other
manufacturer’s equivalents.

• Single shield cable with a minimum of 90%

coverage. Note that a double shielded cable
may be required in high-noise applications.

C4.1

Signal Wiring
Each differential signal should be connected to
a corresponding twisted pair as follows: