Microsonic mic-25/IU/M User Manual

Set the mic-sensor using the teach-in procedure, Instruction manual, Mic ultrasonic sensors with one analogue output

Advertising
background image

89/336/EEC

Product description

 The mic-sensor with one analogue output

measures the distance to an object within
the detection zone contactless. A signal
proportional to distance is created ac-
cording to the adjusted window
margings of the analogue characteristic
curve.

 The sensor automatically detects the load

put to the analogue output and switches
to current output or voltage output
respectively.

 Choosing between rising and falling out-

put characteristic is possible.

 The sensors are adjustable using Teach-in

processes via the Com-channel (Pin 5).

 Using the LinkControl adapter (optional

accessory) all Teach-in and additional sen-
sor parameter settings may be made by a
Windows-Software.

Important instructions for assembly and
application

All employee and plant safety-relevant
measures must be taken prior to assembly,
start-up, or maintenance work (see operation
manual for the entire plant and the operator
instruction of the plant).

The sensors are not considered as safety
equipment and may not be used to ensure
human or machine safety!

The mic-sensors indicate a blind zone, in
which the distance cannot be measured. The
operating range indicates the distance of the
sensor that can be applied with normal re-
flectors with sufficient function reserve.
When using good reflectors, such as a calm
water surface, the sensor can also be used up
to its maximum range. Objects that strongly
absorb (e.g. plastic foam) or diffusely reflect
sound (e.g. pebble stones) can also reduce
the defined operating range.

Assembly instructions

 Assemble the sensor at the installation lo-

cation.

 Plug in the connector cable to the M 12

connector.


Abb. 1: Pin assignment with view onto sensor

plug and colour coding of the
microsonic connection cable

Synchronisation

If the assembly distances shown in Fig. 2 for
two or more sensors are exceeded the inte-
grated synchronisation should be used. Con-
nect Sync/Com-channels (pin 5 at the units re-
ceptable) of all sensors (10 maximum).

1
3

colour

+U

B

-U

B

brown
blue

4
2
5

-

I/U

black
white

Com.

grey

1

5

2

3

4

Abb. 2: Assembly distances

Start-up

mic-sensors are delivered factory made with
the following settings:

 Rising analogue characteristic
 Window margins for the analogue out-

put set to blind zone and operating
range

 Maximum detection range set to maxi-

mum range

Set the parameters of the sensor using the
Teach-in procedure to adjust the analogue
chacteristc curve.

Operation

mic-sensors work maintenance free. Small
amounts of dirt on the surface do not influ-

≤ 0.35 m
≤ 0.40 m

≤ 2.50 m
≤ 2.50 m

≤ 1.10 m
≤ 2.00 m

≤ 2.50 m
≤18.00 m

≤ 4.00 m

≤30.00 m

A

B

ence function. Thick layers of dirt and ca-
ked-on dirt affect sensor function and the-
refore must be removed.

Note

 mic-sensors have internal temperature

compensation. Because the sensors heat
up on their own, the temperature com-
pensation reaches its optimum working
point after approx. 30 minutes of opera-
tion.

 The load put to the analogue output is

detected automatically when turning
supply voltage on.

 If no signal is detected for 20 seconds

during teach-in procedure the made
changes are stored and the sensor re-
turns to normal mode operation.

 You can reset the factory settings at any

time, see »Reset to factory setting«.

Teach-in switched output

Instruction manual

mic Ultrasonic Sensors with

one analogue output

mic-25/IU/M
mic-35/IU/M
mic-130/IU/M
mic-340/IU/M
mic-600/IU/M

Additional settings

Set the mic-sensor using the Teach-in procedure

Set window margins

Place object

at position 

Connect Com for about

3 s with + U

B

Place object

at position 

Connect Com for about

1 s with + U

B

Set rising / falling output

characteristic curve

Connect Com for about

13 s with + U

B

To change output

function

connect Com for about

10 s with + U

B

Change Teach-in / Syn-

chronisation

Turn supply voltage OFF

Reset to factory setting

Turn supply voltage OFF

While Com is connected

to - U

B

turn supply

voltage ON

Keep Com connected to

- U

B

for about 3 s

To change operation

mode (Teach-in and Syn-

chronisation)
connect Com

for about 1 s to - U

B

While Com is connected

to - U

B

turn supply

voltage ON

Keep Com connected to

- U

B

for about 13 s

Disconnect

Com from - U

B

before swit-

ching-off

supply voltage

Normal operation mode

Wait for 10 s

Wait for 20 s

Normal operation mode



Tech-in analogue output

Further Settings