Seaward 100 PrimeTest User Manual
Page 19
Advertising

PrimeTest 100
Operating Instructions
4 Specifi cation
Earth Continuity
Accuracy*
± (5% + 2 digits)
Test current
200mA minimum
Test voltage
9V nominal
Insulation resistance
Accuracy
± (5% + 2 digits)
Test
voltage
500V
Test current
>1mA into 500kΩ
Test current
<2mA into 2kΩ
Leakage Current
Accuracy
± (5% + 2 digits)
Test voltage 40V rms, 50Hz AC
Test current <5mA into 2kΩ
Cord Test
Earth continuity, insulation resistance as above.
Check for Live and Neutral open circuit, short
circuit or reversed polarity.
*When used with Seaward test lead, Part Number 347A002
Factory Set Pass/Fail limits
Class I
0.2 ohms
1.0Mohm
0.75mA
Class II
N/A
2.0Mohm
0.25mA
Cord
0.2 ohm
2.0Mohm
N/A
Earth
Continuity
Insulation
Resistance
Leakage
17