eFlo Electronic Flow Meter Operations Manual
Super Systems Inc.
Page 16 of 25
5.
Using the manometer, measure the actual differential pressure at the selected flow
rate. Compare the differential pressure to the differential pressure shown on the
Calibration Report.
NOTE: Typically, a 0.5”wc deviation equates to 5% flow error.
6.
If the actual differential pressure is within acceptable limits (±4% of the full scale value),
no action is necessary. If the actual differential pressure is not within acceptable limits,
adjust the zero/span potentiometers inside the differential pressure transducer. The
pressure transducer is located inside the black enclosure on top of the unit.