Troubleshooting – Tecfluid COVOL Series User Manual

Page 23

Advertising
background image

23

10

TROUBLESHOOTING

Problem

Possible cause

Solution

The associated
electronics does not
totalize

Blockage or friction of the piston
by particles

Install a filter according to
specification (maximum mesh
size between 0.1 y 0.2 mm²).

Damaged reed group

Change the reed group

Disconnected cables between the
COVOL and the associated
electronics

Check the cable connection

The associated
electronics totalize less
than the actual volume

Operating liquid with properties
very different from water at 20ºC
(calibration liquid)

Correct the i/l factor

Blockage or friction of the piston
by particles

Install a filter according to
specification (maximum mesh
size between 0.1 y 0.2 mm²).

The associated
electronics totalizes
more than the actual
volume

Operating liquid with properties
very different from water at 20ºC
(calibration liquid)

Presence of air pockets

Control the minimum tank level.
Provide an air degasser
upstream of the counter if air
intakes can be expected.

Excessive emptying of storage
tanks

The flow rate indicated by
the associated equipment
is unstable

The pipe is not completely full

Make sure that the pipe is
completely full, for example,
installing the counter in a vertical
pipe with upwards flow

Correct the i/l factor