3.

INSTALLATION OF THE INTERCONNECTION CABLE

The following points must be taken into account:

•

The cable should be installed in a conduit or should be securely fixed, given that
movements of the cable can induce reading errors.

•

The placement of the cable should be as far away as possible from sources of electrical
noise switching gear and electrical machines.

•

The maximum length of the cable L depends on the fluid conductivity. The minimum
conductivity of 5

µS/cm is for a length of up to 5 m.

4. CABLE

CONNECTION