Watson-Marlow 313F User Manual

Page 8

Advertising
background image

8

313T major components

1

PCB cover

2

"Off time" knob

3

Pushbutton switch

4

Slider switch

5

"On time" knob

6

25 Din socket

Care and maintenance

The only scheduled maintenance of the 313F, 313S and 313U is to inspect the motor brushes and to replace them before
their length is less than 6mm 1/4". The life of the brushes will depend on the duty of the pump, which is expected to be
at least 7,500 hours at maximum speed. When the pump needs cleaning, remove the pumphead and use a mild solution
of detergent in water. Do not use strong solvents.

If the gearbox is dismantled, it should be filled with a good quality grease such as Castrol MS3.

Specification

313F/D nominal rotor speeds

5/50rpm, 200/400rpm

313U/D, 313S/D maximum rotor speeds

50rpm and 400rpm

Operating voltages and frequencies

110-240V 50/60Hz

313 S/D, 313U/D speed control ratio

20:1

Power rating

100VA

Operating temperature

5C to 40C

Storage temperature range

-40C to 70C

Weight

5.35kg, 12Ib

Noise

<70dBA at 1m

Standards

IEC 335-1, EN60529 (IP31)

Machinery Directive:98/37/EC EN60204-1

Low Voltage Directive:73/23/EEC EN61010-1

EMC Directive:89/336/EEC EN50081-1/ EN50082-1