8 PROGRAMMING

8.1 Programming

principle


In the following, a parameter will be the value of a single function, i.e. volume, tube diameter or
velocity. Programming is done from a master controller like MC12P from Flexicon.

A program will be a complete set of parameters which together will constitute the PD22 work
instructions.

The actual programming will be made on the MC12 and reference is made to the MC12
manual.

8.2 PD22

parameters


1. Volume
Value: ml
Range: 10 - 9999 ml
PD22 can dispense from 10 to 9999 ml, but for volumes exceeding approx. 1000 ml, the
capacity will drop significantly.

If the volume is changed, a recalibration should be made.

2. Tubes
Value: Inside diameter (i.d.) in mm
Range: 3.0 - 5.0 - 6.5 - 8.5 - 10.0 - 12.5
The tubes can be measured with the supplied tube gauge.

If the tube is changed, a new calibration must always be made.

3. Velocity
Value: Revolutions per minute (rpm)
Range: 30 - 250 rpm

If the velocity is changed, a recalibration should be made.

4. Acceleration/deceleration
Value: An integral number
Range: 1 - 100
The acceleration and the deceleration will always be the same. The lowest value (1) will give
the slowest acceleration, and the highest value (100) will give the fastest acceleration.

If the acceleration is changed, a recalibration should be made.

5. Reversing (back suction)
Value: An integral number
Range: 0 - 10