Reference Manual

00809-0100-4811, Rev CA

February 2006

Rosemount 3300 Series

5-2

ERROR MESSAGES

The display can also be used for presentation of software errors. The upper
row shows error codes and the lower row shows 'ERROR'.

Figure 5-2. Presentation of error
messages

The following errors can be displayed:

See also “Errors” on page 6-22.

ALARM AND WRITE
PROTECTION

When mounting the Integral Display panel it is important that the Alarm and
Write Protection switches on the transmitter mother board are correctly set.
Make sure that the Alarm switch is in the HIGH position and the Write
Protection switch is in the OFF position, see Figure 5-3. See also Section 3:
Before You Install
for more information.

Figure 5-3. Alarm and Write
Protection switches.

Once the mother board positions are set, then the display positions become
the master.

Code

Error

CNFIG

Invalid Configuration

00001

Ram Failure

00002

ROM Checksum

00006

Waveform Acquisition Failure

00007

EEprom Factory Checksum

00008

EEprom User Checksum

00010

Software Error

00013

Probe Failure

Error code