Sorting and tolerance checking – HEIDENHAIN ND 282 B User Manual
Page 12

12
Sorting and Tolerance Checking
In the sorting and tolerance checking mode, the display unit
compares the displayed value with the programmed upper
and lower sorting limits. The sorting and tolerance checking
mode is enabled and disabled with operating parameter
P17.
Entering sorting limits
Sorting limits are entered in operating parameters P18 and
P19 (see “Operating Parameters”).
Sorting signals
The indicators and switching outputs at D-sub connection
EXT (see section on X41) sort the display value into one of
three classes.
Display
Meaning
=
Measured value is within sorting limits
<
Measured value is smaller than lower sorting
limit
>
Measured value is greater than upper sorting
limit
Operating parameters for sorting and tolerance
c h e c k i n g
P17 CLASS.
Sorting ON/OFF
P18 L.CLASS.
Lower sorting limit
P19 U.CLASS.
Upper sorting limit
Example: Upper sorting limit = 26.02 mm
Lower sorting limit = 26.00 mm
Part 1
Part 2
Part 3
Sorting and Tolerance Checking
Reject
Accept
Rework