Bcd data interface (x33) – HEIDENHAIN ND 282 B User Manual
Page 28

28
Triggering the “Contact” and “Pulse” inputs at D-sub
connection EXT (X41)
BCD Data Interface (X33)
Pin 1(0V)
Pin 23
Pin 22
Pin 1(0V)
EXT(X41)
EXT(X41)
Signal run times
The times in the following table are guide lines.
If you are working with slow data output and are using display
functions at the same time (e.g. measurement series or inch display),
the actual signal run times can be twice as long as given here.
Concurrent Data Output (P55 ACTL)
Operating
mode
P53
Latch rate
Data output after
fast
FAST
P54
Value from P54 / 2
slow
SLOW
t
£ 30 ms
t
£ 8 ms
Data output after external latching
(P55 STOP or HOLD)
Operating P53
Minimum
Measured value
mode
impulse length
latched after
Pulse/
Contact BCD
Pulse
Contact
BCD
fast
FAST
3 µs
7 ms
0.3 µs 1.1 µs
4.8 ms
slow
SLOW
t
³ 8 ms t ³ 13 ms 0.3 µs 1.1 µs
4.8 ms
Data output
Relatch after
Pulse/
Contact
BCD
fast
FAST
£ 0.3 µs after internal
3 µs
7 ms
latching
slow
SLOW
£ 7.5 ms after internal
3 µs
7 ms
latching