Sorting and Tolerance Checking
In the sorting and tolerance checking mode, the display unit
compares the displayed value with the programmed upper and
lower sorting limits. The sorting and tolerance checking mode is
enabled and disabled with operating parameter P17.
Entering sorting limits
Sorting limits are entered in operating parameters P18 and P19
(see “Operating Parameters”).
Sorting signals
The indicators and switching outputs at D-sub connection EXT
(see section on X41) sort the display value into one of three
classes.
Display
Meaning
=
Measured value is within sorting limits
<
Measured value is smaller than lower sorting limit
>
Measured value is greater than upper sorting limit
Operating parameters for sorting and tolerance checking
P17 SORT
Sorting ON/OFF
P18 L.LIMIT
Lower sorting limit
P19 U.LIMIT
Upper sorting limit
Example: Upper sorting limit = 26.02 mm
Lower sorting limit = 26.00 mm
Part 1
Part 2
Part 3
Sorting and Tolerance Checking
Reject
Accept
Rework