4 burner internal components, A. disassembly of burner/thermal block, 6 burner/thermal block disassembly – Emerson Process Management 760004-A User Manual

Page 46: Model nga 2000 hfid

Advertising
background image

Instruction Manual

760004-A
February 2002

4-8 Maintenance and Service

Rosemount Analytical Inc. A Division of Emerson Process Management

Model NGA 2000 HFID

4-4 BURNER INTERNAL COMPONENTS

WARNING

BURNER CONTAMINATION

Do not handle internal parts of the burner
with bare hands. All tools used for main-
tenance must be free of contaminates.

a. Disassembly of Burner/Thermal Block

1. Remove oven from analyzer module

per Section 4-2a.

2. Remove burner/thermal block from

oven per Section 4-2b.

3. Refer to Figure 4-6. Disconnect sam-

ple capillary nut at base of burner.

4. Remove screw securing thermal block

to burner.

5. Carefully pull burner away from ther-

mal block.

Figure 4-6. Burner/Thermal Block Disassembly

Sample capillary

Sample capillary
input

Burner

Thermal Block

Advertising