B. replacing burner jets, 7 burner disassembly, Model nga 2000 hfid – Emerson Process Management 760004-A User Manual

Page 47

Advertising
background image

Instruction Manual

760004-A

February 2002

Rosemount Analytical Inc. A Division of Emerson Process Management

Maintenance and Service 4-9

Model NGA 2000 HFID

Burner Cap

Burner Cap
Retainer

Gasket
656931

Burner Base

Burner Base

Combustion
Chamber

A

B

O-Ring
904373

b. Replacing Burner Jets

Disassemble the burner only if contami-
nants are evident. Combustion products
or other contaminates which accumulate
inside the burner may form electrical
leakage paths between the collector and
the burner contact, resulting in noisy
readings.

If the analyzer module is to be operated at
the highest sensitivity, traces of such
contaminates can cause erroneous read-
ings. For best performance, replace the
burner jet follows:

WARNING

BURNER CONTAMINATION

Do not handle internal parts of the burner
with bare hands. All tools used for main-
tenance must be free of contaminates.

1. Remove oven from analyzer module

per Section 4-2a.

2. Remove burner/thermal block from

oven per Section 4-2b.

3. Remove thermal block from burner

per Section 4-4a.

4. Refer to Figure 4-7A. Remove

screws (2) holding burner cap re-
tainer, remove retainer.

5. Holding burner base, lift burner cap

off of assembly, set aside, remove
gasket.

6. Refer to Figure 4-7B. Holding burner

base, lift combustion chamber off, set
aside.

Figure 4-7. Burner Disassembly