Instruction Manual

IB-106-300NH Rev. 4.2

July 2002

Rosemount Analytical Inc. A Division of Emerson Process Management

Installation 2-11

World Class 3000

NOTE

General Wiring Recommendations

To maintain CE compliance and ensure proper EMC performance, all signal wires to the Inter-
connect Board, with the exception of the probe cable, should be looped through the ferrite
beads provided as shown in Figure 2-6 (P
/N 1L04253H01). Signal wires may be grouped to-
gether and looped through before exiting the enclosure. Ferrite beads should be placed as
close as possible to the exit point. Ferrite beads are provided for European Common Market
applications only.

Figure 2-6. Signal Wire Routing