World class 3000, Instruction manual, Intelligent field transmitter ift 3000 – Emerson Process Management 3000 User Manual

Page 51

Advertising
background image

Instruction Manual

IB-106-300NH Rev. 4.2

July 2002

Rosemount Analytical Inc. A Division of Emerson Process Management

Installation 2-19

World Class 3000

JM7

I

V

MPS TERMINATION BOARD

MPS 3000 MULTIPROBE CALIBRATION GAS SEQUENCER (OPTIONAL)

L
E

N

LINE

VOLTAGE

J13 J14

J15

J16

J17 J18

J12

CAL

RET

HI

GAS

IN

CAL

NO

GAS

CAL

RET

HI

GAS

IN

CAL

NO

GAS

CAL

RET

HI

GAS

IN

CAL

NO

GAS

CAL

RET

HI

GAS

IN

CAL

NO

GAS

LOW

GAS

LOW

GAS

LOW

GAS

LOW

GAS

NC C NO NC C NO NC C NO NC C NO

L

N

L

N

LINE OUT

LINE IN

J10

J11

PROBE 1

PROBE 2

PROBE 3

PROBE 4

PROBE 1

PROBE 2

PROBE 3

PROBE 4

PROBE

1

SOLENOID

PROBE

2

SOLENOID

PROBE

3

SOLENOID

PROBE

4

SOLENOID

HIGH

GAS

SOLENOID

LOW

GAS

SOLE

NOID

PRESSURE SWITCH

SHIELD

SHIELD
RELAY –
RELAY +

TRIAC +

CAL RET

SHIELD

NO GAS

AD590 –

LO GAS

AD590 +

HI GAS

SHIELD

IN CAL

TRIAC –

SHIELD

SHIELD

STACK TC –
STACK TC +

PROBE MV –
PROBE MV +

PROBE TC +

PROBE TC –

INTELLIGENT FIELD

TRANSMITTER IFT 3000

ALWAYS DISCONNECT LINE VOLTAGE

FROM INTELLIGENT FIELD TRANSMITTER

BEFORE CHANGING JUMPERS.

LINE

VOLTAGE

SECTION

JUMPER

(INSTALL)

PROBE HEATER

VOLTAGE SECTION

JUMPER

(INSTALL)

100 V.A.C.
120 V.A.C.

220 V.A.C.

200 V.A.C.

240 V.A.C.

J1

J5

J6

J1

JM1

JM6

J2

J3

J4

J5

J6

J7

J8

J9

3D39122G REV

POWER SUPPLY BOARD

3D39513G

MICROPROCESSOR

BOARD

3D39120G REV

INTERCONNECT BOARD

JM3, JM7, JM2
JM8, JM7, JM1

JM6, JM5, JM2

JM4, JM5, JM2

JM6, JM5, JM1

NOT USED

REMOVE

JM9, JM10

JUMPER CONFIGURATION

E

N

L

LINE

VOLTAGE

NOT USED

5 CONDUCTOR SHIELDED CABLE

PER PROBE #16 AWG BY CUSTOMER

CURRENT/VOLTAGE

SELECTOR SWITCH

27270012

A

B

Figure 2-14. Wiring Layout for Complete IFT 3000 System with HPS (Sheet 2 of 2)

2

Advertising