IB-106-300NH Rev. 4.2

July 2002

Rosemount Analytical Inc. A Division of Emerson Process Management

Installation 2-21

World Class 3000

1

2

NOTES:

100 V.A.C. OPERATION REQUIRES TRANSFORMER PART
NUMBER 1M02961G02.

REFER TO TABLE 3-5 FOR PROPER SET POINT SELECTION.

1

2

0310122

Figure 2-16. Jumper Selection Label

NOTE

Before supplying power to the heater
power supply, verify that jumpers JM3,
JM6 are removed and JM7 is installed.
If relay wire (Figure 2-14, note 1) i
s in-
stalled, JM2 must be removed from
HPS Mother Board (Figure 2-17).

4. Before supplying power to the heater

power supply, verify that the jumpers
on the mother board, Figure 2-17, are
properly configured. Jumpers JM3 and
JM6 should be removed and JM7
should be installed.

Additionally, make sure that the proper
jumper for your line voltage is installed,
Figure 2-16. If relay wire (Figure 2-14,
note 1) is not installed, JM 2 should be
installed on the HPS Mother Board
(Figure 2-17).

NOTE

Refer to Figure 2-9 and Figure 2-11 for
proper IFT jumper configuration. IFT
microprocessor and interconnect
board jumper configurations must be
set correctly in order for HPS to work
properly.

29850001

NOTE: ALL FUSES ARE 5A @ 250 VAC,

ANTI-SURGE, CASE SIZE

5 X 20 MM, TYPE T TO IEC127,

SCHURTER.

3D3

080G

REV

JM1

JM2
JM4

JM5

JM7

FUSE

FUSES

Figure 2-17. Jumpers on HPS Mother Board

2