Model IFT9701 Transmitter Instruction Manual

Configuration with ProLink II Software

continued

7.2

Configuration parameters

Use the Configuration window to perform the following tasks:

•

Assigning a HART tag to the flowmeter

•

Changing process variable measurement:

-

Changing measurement units for the mass flow or volume flow rate

-

Changing the flow cutoff value

-

Changing the internal damping value

-

Changing the flow direction parameter

•

Setting range values for the milliamp output

•

Scaling the pulse output

7.2.1

HART tag

The HART tag consists of up to eight characters that identify the flowmeter when it communicates
with other devices in a HART multidrop network.

To assign a HART tag to a flowmeter:

1. From the ProLink menu, click on Configuration.

2. In the Configuration window, click on the Device tab.

3. When the Device panel appears (as depicted in Figure 7-1), enter the desired tag of up to eight

characters, then click on OK. The tag can include spaces and periods.