Sealey TA090 User Manual

Page 13

Advertising
background image

NOTE:

14.9.5 Apply the probe to test sample several times.

14.9.6 The thickness of the foil should be roughly equivalent to the estimated coating thickness.

14.9.7 Even while a series of measurements is being taken, foil calibration can be carried as often as

necessary. The old calibration will be overwritten; the ZERO calibration remains in memory until a

new zero point calibration is made.

14.9.8 See one-point calibration for more information.

14.10 Shot-blasted surfaces

The physical nature of shot-blased surfaces results in coating thickness reading that are too high. The

mean thickness over the peaks can be determined as follows (The statistics program is not of great

benefit in this procedure):

Method A:

14.10.1 The gauge should be calibrated according to 14.8 or 14.9. Use a smooth calibration sample

with the same curvature radius and the same substrate as the later measuring sample.

14.10.2 Now take approx. 10 readings on the uncoated, shot-blased sample to produce the mean

value Xo

14.10.3 After this take approx. 10 further readings on coated, shot blasted test sample to produce the

mean value Xm 4. The difference between the two mean values is the mean coating thickness Xeff

over the peaks. The greater standard deviations of the two values Xm and Xo should also be taken

into consideration: Xeff=(Xm-Xo)±S

Method B:

14.10.4 Carry out a zero calibration of 10 readings on a shot-blased, uncoated sample. Then carry

out a foil calibration on the uncoated substrate. The foil set should consist of a number of individual

foils of max. 50 microns thickness each and should roughly correspond to the estimated coating

thickness.

14.10.5 The coating thickness can be read directly from display and should be averaged from 5 to 10

single measurements. The statistics function is useful here.

Method C:

This method also gives reliable results. Simply follow the two-point calibration method using two foils

as described in section 14.9. For a maximum approach to the respective nature of surface, the foil

value can be reached by using several foils – 50um each. The mean coating thickness should be

calculated from 5 to 10 readings. The statistics program is very useful here.

NOTE:

For coatings thicker than 300 um, the influence of roughness generally is of no importance and it will

not be necessary to apply above calibration methods.

14.11 General Remarks on Measurement

After careful calibration has been made, all subsequent measurements will lie within the guaranteed

measuring tolerance. Strong magnetic fields near generators or live rails with strong currents can

affect the reading.

When using the statistics program for obtaining a mean value it is advisable to place the probe

several times at a typical measuring spot. Any false readings can be removed immediately Via the

MENU system.

The final reading derives from the statistical calculation and from the guaranteed tolerance levels of

the gauge.

Coating Thickness D =X±s±u

Example: Readings: 150um, 156um, 153um

Mean value: X=153um

Standard deviation: s=±3um

Measuring uncertainty: u=±(1% of reading + 1um)

D= 153±3±(1,53um+1um)

=153±5,5um

MAX: maximum reading

MIN: minimum reading

Original Language Version

TA090 Issue No.1 25/10/12

© Jack Sealey Limited

Advertising