Specification 4. operation – Sealey TA101 User Manual

Page 2

Advertising
background image

Range Resolution Accuracy

326mV 0.1mV

± (0.5% + 5)

3.26V 1mV

32.6V 10mV

± (0.8% + 5)

326V 0.1V

1000V

1V

± (1.0% + 8)

DC Voltage (Autorange)

Input Impedance: 10MΩ (for 326mV range: >100MΩ)

Overload Protection: DCV1000V; ACV750V.

Range Resolution Accuracy

3.26V

1mV

± (0.8% + 5) (40~200Hz)

32.6V

10mV

± (1.2% + 7) (200~400Hz)

326V

0.1V

± (1.0% + 5) (40~200Hz)

750V

1V

± (1.5% + 7) (200~400Hz)

AC Voltage (Autorange)

Range Resolution Accuracy

326µA 0.1µA

3260µA 1µA

± (1.5% + 5) (40~200Hz)

32.6mA 10µA

± (1.8% + 7) (200~400Hz)

326mA 0.1mA

10A

10mA

± (3.0% + 5)

AC Current (µA and mA are Autorange)

Range Resolution Accuracy

326µA 0.1µA

3260µA 1µA

± (1.2% + 5)

32.6mA 10µA

326mA 0.1mA

10A

10mA

± (2.0% + 5)

DC Current (µA and mA are Autorange)

Overload Protection: F 500mA L 250V

(for 10A range: F10A L 250V, Maximum measurement duration

<10 secs. and interval >15 mins).

Range Resolution Accuracy

326Ω

0.1Ω

± (1.0% + 8)

3.26kΩ

1Ω

32.6kΩ

10Ω

± (1.0% + 5)

326kΩ

0.1

kΩ

3.26MΩ

1kΩ

32.6MΩ

10kΩ

± (3.0% + 7)

Resistance (Autorange)

Frequency (Manual Range)

Range Resolution Accuracy

-20°C~750°C 1°C -20~0°C(-4°F~32°F) ±(6.0% +5)

0~400°C(32°F~752°F) ±(1.5% +5)

-4°F~1400°F 1°F 400~750°C(752°F~1400°F) ±(1.8%+5)

Temperature

Overload Protection: F 500mA L 250V

(for 10A range: F10A L 250V, Maximum measurement duration

<10 secs. and interval >15 mins). Frequency: 40Hz to 400Hz.

Range Scope (RPM) Resolution Accuracy

RPM

0~3260

1 RPM

± (2.5% + 5)

10 x RPM 10 x (0~3260) 10 RPM

Tachometer

Description

Test condition

Cylinder Range

Resolution Accuracy

2CYL

0 ~180.0°

3CYL

0 ~120.0°

4CYL

0 ~ 90.0° 0.1°

± (2.5% + 5)

5CYL

0 ~ 72.0°

6CYL

0 ~ 60.0°

8CYL

0 ~ 45.0°

Dwell Angle (Manual Range)

Range Resolution

Accuracy

320Hz 0.1Hz

3200Hz 1Hz

± (2.5% + 5)

32kHz 0.1kHz

Diode test

Range Scope (%) Resolution Accuracy

Duty 0~99.9 0.1

± (2.0% + 5)

Duty Cycle test

Audible Continuity Test

Input Impedance: 10MΩ

Frequency Range: 40Hz to 400Hz

Overload Protection: DC 1000V; AC 750V.

If the resistance of the circuit

under test is lower than 50Ω,

an audible warning will sound.

Open circuit voltage is

approximately 3V.

Description

Test condition

The approximate forward

voltage of the diode under test

will be displayed on the LCD.

The forward DC current is

approximately 1mA, the

reversed DC voltage is

approximately 3V.

WARNING! Ensure that you read, understand and apply the safety and operational instructions before connecting the analyser. Only when you are sure

that you understand the procedures is it safe to proceed with testing.

WARNING! Risk of electrocution. High voltage circuits, both AC and DC are very dangerous and should be measured with great care.

Operating temperature range 0

°C to 40°C.

Remember to turn off analyser when measurement is completed.

Note: If “OL” appears in the display during a measurement, the value exceeds the range you have selected. Select a higher range.

Note: On some low AC and DC ranges, with the test leads not connected to a device, the display may show a random fluctuating reading. This is

normal and is caused by the high input sensitivity. The reading will stabilise and give a correct measurement when connected to a circuit.

4.1.

Data Hold Button

4.1.1. The data hold function allows the analyser to freeze a measurement reading for later reference.

4.1.2. Press the data hold button once to freeze the reading in the display. The indicator “H” will appear in the display.

4.1.3. Press the data hold button again to return to normal operation.

4.2.

AC/DC Voltage Measurement

4.2.1.

Insert the black test lead into the negative COM socket and the red test lead into the positive

V

RPM

socket.

4.2.2. Turn the function dial to the voltage setting and press the function selection button to select AC or DC as required.

4.2.3. Connect the test leads to the circuit under test and read the voltage on the display.

4.2.4. For DC voltage, the polarity of the red test lead will be displayed as well as the value.

4.3.

DC Current Measurement

WARNING! Do not make current measurements at 10A for longer than 10 seconds in every 15 minutes. Exceeding this may cause damage to

the analyser and test leads.

4.3.1. Insert the black test lead into the negative

COM socket, and the red test lead into the positive µAmA socket (or the positive 10A socket for currents from

326mA

to

10A).

4.3.2. Turn the function dial to the appropriate setting as required (see ranges above). If the current range to be measured is not known, set to the highest

range setting, and then select to the correct range when the first reading is taken, until satisfactory resolution is obtained. Press the function selection

button to select AC or DC as required.

4.3.3. Connect the test leads in series with the circuit under test and read the current on the display.

4.3.4. For DC current, the polarity of the red test lead will be displayed as well as the value.

Original Language Version

3. SPECIFICATION

4. OPERATION

© Jack Sealey Limited

TA101 Issue No.3(L) - 16/06/14