Precision Digital PD6300 User Manual

Page 43

Advertising
background image

Model PD6300 Pulse Input Rate/Totalizer

Instruction Manual

43

K-Factor Calibration (Factor)

The meter may be calibrated using the K-Factor function. Most flowmeter
manufacturers provide this information with the device. Enter the
K-Factor

(Factor) menu and select the decimal point with highest reso-

lution possible and program the k-factor value (i.e. pulses/gal). The me-
ter will automatically calculate the flow rate using the k-factor and the
time base selected.

ddd.ddd

factor

Factor

prog

d.ddddd

factor

1.8925

6

factor

Multi-Point Calibration & Scaling

The meter is set up at the factory for 2-point linear calibration. The num-
ber of points for multi-point calibration/scaling is set up in the Advanced
Features

menu. Up to 32 linearization points may be selected. See page

76 for details.

Advertising
This manual is related to the following products: