Maintenance, Changing the battery – Amprobe AC50A User Manual

Page 11

Advertising
background image

Maintenance

Provided it is used in accordance with the users manual, the instrument
needs no special maintenance.

Cleaning
If the instrument is dirty after daily usage, it is advised to clean it by using
a humid cloth and a mild household detergent.

Prior to cleaning, ensure that instrument is switched off and dis-

connected from external voltage supply and any other instruments
connected (such as UUT, control instruments, etc.).
Never use acid detergents or dissolvants for cleaning.

Changing the battery

�

Prior to storage battery replacement, disconnect the instrument
from any circuits.

Only use batteries as described in the technical data section.

• If the symbol for Low-battery appears in the upper left corner of the

display, the

battery must be changed.

This is carried out as follows:
1) Separate the Amprobe Current Clamp AC50A from any circuit, and

remove the test leads.

2) Switch the instrument off
3) Open the housing by removing the 3 screws on the rear face.
4) Remove the old batteries
5) Fit the new batteries (2 type 1.5 V IEC LR6), taking care that the polarity

is correct. Make sure that no wires are trapped between the 2 halves
of the housing, and close it again.

6) The instrument is now ready for further use.

Please consider your environment when you dispose of your one-way
batteries or accumulators. They belong in a rubbish dump for hazardous
waste. In most cases, the batteries can be returned to their point of
sale.

�

Please, comply with the respective valid regulation regarding the return,
recycling and disposal of used batteries and accumulators.

If an instrument is not used over an extended time period, the batteries
must be removed. Should the instrument be contaminated by leaking
battery cells, the instrument has to be returned for cleaning and inspec-
tion to the factory.

9

Advertising