Amprobe ACD-14-PLUS Clamp-On-Multimeters User Manual

Page 15

Advertising
background image

12

ACA Current (Clamp-on 50Hz / 60Hz)

Range

Accuracy

1) 2) 3)

40.00, 400.0, 600 A

±( 1.5% rdg + 8 digits)

Overload Protections:

ACA Clamp-on jaws: 600 A rms continuous

ACD-14 Plus:

Average Sensing

ACD-14 TRMS Plus:

True RMS sensing - 10 % to 100 % of range

1) Max Induced error from adjacent current carrying conductor: 0.05 A
2) Specified accuracy is from 1% rdg to 100% rdg of range and for measurements

made at the jaw center. When the conductor is not positioned at the jaw center,

position errors introduced are: Add 2% rdg to specified accuracy for measurements

made BEYOND jaw marking lines (toward jaw opening)

3) Add 8 digits to specified accuracy @ reading < 10% rdg of range

Frequency

Function

Sensitivity (Sine RMS)

Range

Accuracy

400.0 mVac

350mV

10 Hz ~ 2 kHz

±( 0.5% rdg + 4 digits)

4.000 Vac

1V

5 Hz ~ 5 kHz

±( 0.5% rdg + 4 digits)

4.000, 40.00 Vac

32V

5 Hz ~ 100 kHz

±( 0.5% rdg + 4 digits)

400.0 Vac

90V

5 Hz ~ 10 kHz

±( 0.5% rdg + 4 digits)

600 Vac

500V

5 Hz ~ 5 kHz

±( 0.5% rdg + 4 digits)

Display counts: 5000
Resolution: 0.001Hz
Transient protection : VAC input jacks : 6.5kV (1.2/50μs surge)

Ohms

Range

Accuracy

400.0 Ω

±( 0.8% rdg + 8 digits)

4.000, 40.00, 400.0 kΩ

±( 0.6% rdg + 4 digits)

4.000 MΩ

±( 1.0% rdg + 4 digits)

40.00 MΩ

±( 2.0% rdg + 4 digits)

Open Circuit Voltage : 0.4 VDC typical
Transient protection : 6.5 kV (1.2/50μs surge)