13

Capacitance

Range

1)

Accuracy

2) 3)

500.0nF, 5.000μF, 50.00μF,
500.0μF, 3000μF

±( 3.5% rdg + 6 digits)

1) Additional 50.00nF range accuracy is not specified
2) Accuracies with film capacitor or better
3) Specified with battery voltage above 2.8V (approximately half full battery).
Accuracy decreases gradually to 12% rdg at low battery warning voltage of
approximately 2.4V
Transient protection: 6.5 kV (1.2/50 μs surge)

DC μA

Range

Accuracy

Burden Voltage

400.0μA

±( 2.0% rdg + 4 digits)

2.8mV/μA

2000μA

±( 1.2% rdg + 3 digits)

2.8mV/μA

Transient protection: 6.5 kV (1.2/50 μs surge)

AC μA (50Hz ~ 500Hz)

Range

Accuracy

Burden Voltage

400.0μA

±( 2.0% rdg + 5 digits)

2.8mV/μA

2000μA

±( 1.2% rdg + 5 digits)

2.8mV/μA

Transient protection: 6.5 kV (1.2/50 μs surge)

Type-K Temperature

Range

Accuracy

-20 °C ~ 300 °C

±( 2% rdg + 3 °C)

301 °C ~ 537 °C

±( 3% rdg + 3 °C)

-4 °F ~ 572 °F

±( 2% rdg + 6 °F)

573 °F ~ 999 °F

±( 3% rdg + 6 °F)

Type-K thermocouple range & accuracy not included